Stand-Alone Anterior Cervical Discectomy and Fusion Using an Additive Manufactured Individualized Bioactive Porous Titanium Implant without Bone Graft: Results of a Prospective Clinical Trial
Article information
Abstract
The purpose of this study was to introduce our patient-specific bioactive porous titanium implant manufactured using selective laser melting (SLM) and to establish the efficacy and safety of the implant for stand-alone anterior cervical discectomy and fusion (ACDF) based on a prospective clinical trial. We designed a customized ACDF implant using patient-specific data and manufactured the implant using SLM. We produced a bioactive surface through a specific chemical and thermal treatment. Using this implant, we surgically treated four patients with cervical degenerative disc disease and evaluated the clinical and radiological results. We achieved successful bony union in all but one patient without autologous bone grafting within 1 year. We observed no implant subsidence during the follow-up period, and all clinical parameters improved significantly after surgery, with no reported implant-related adverse effects. Our customized bioactive porous titanium implant is a safe and promising implant for stand-alone ACDF.
Introduction
Anterior cervical discectomy and fusion (ACDF) with autologous iliac bone graft is the gold standard for treating cervical degenerative disc disease [1,2]. However, autologous bone grafts obtained from the anterior iliac crest are associated with several complications, including donor site morbidities, graft collapse, and fusion failure [3-5]. Several types of stand-alone interbody fusion cages have recently been developed; however, cage subsidence and nonunion are still major concerns related to the mismatch of the shape and mechanical properties of cages with vertebral endplates [6].
Porous titanium and its alloys have the greatest potential as orthopedic implants for load-bearing applications, and their interconnected pore structure can facilitate bone ingrowth. Although there are various methods for producing porous titanium, including particle sintering, spraying a plasma of titanium powder, and compressing and sintering of titanium fibers [7-9], none of these methods can completely control the shape, size, or porosity of the pore.
Rapid prototyping technologies (previously used only for surgical planning) have been demonstrated to be effective rapid manufacturing techniques through additive manufacturing (AM). Selective laser melting (SLM) using titanium powder is an AM technology that can produce porous titanium implants with a particular shape and pore structure [10]. Previous studies have suggested that porous titanium fabricated by AM accelerates osteogenesis [11-13]. The geometric freedom allows for the manufacture of implants with specific structures and mechanical properties, making AM technology especially useful for manufacturing patient-specific devices and implants.
In the current study, we introduce the individualized bioactive porous titanium implant and report the clinical results of this prospective clinical trial.
Technical Note
1. Designing methods for the customized implant
We obtained preoperative computed tomography (CT) scans of the affected cervical spine (1-mm-thick slice) and transferred the CT data into commercially available image processing software (VG Studio; Volume Graphics, Heidelberg, Germany) to reconstruct a three-dimensional (3D) model of the bone structures in a multiplanar reconstruction (MPR) view. The 3D vertebral bone structures were reconstructed by a segmentation process in which the optimal threshold value was selected by visual inspection (Fig. 1). The 3D data of the bone structures were converted to stereolithography (STL) format and transferred to 3D-CAD (computer-aided design) software (FreeForm; SensAble Technologies, Woburn, MA, USA) for implant design. For the basic form of the porous structure, we adapted a diamond crystal lattice, in which each atom is tetrahedron-like and surrounded by four other atoms (Fig. 2). We designed the manufactured porous titanium implants with a porosity of 65% and pore sizes of 600 μm [14]. The STL format files were transferred to the SLM machine (EOSINT M270; Electro Optical Systems, Munich, Germany).
2. Fabrication of the individualized implant and bony models
Individualized titanium implants were fabricated using an SLM method with commercially pure titanium powder (grade 2, particle size <45 µm; Osaka Titanium Technologies Co. Ltd., Osaka, Japan). Titanium powder was melted using an ytterbium fiber laser beam in an argon gas atmosphere. The selected slice of the product was then solidified. The top of the previously melted surface was recoated with a 30-µm-thick layer. Subsequently, selective irradiation was performed using the laser beam. We repeated these steps until the final geometry was achieved. The SLM process parameters were as follows: laser power, 120 W; scanning speed, 200 mm/sec; hatch spacing, 120 µm; and hatch offset, 20 µm. We manually evaluated the adaptation and stability of the customized implant using the patient-specific bone model (Z-Printer 450; Z Corp., Burlington, MA, USA).
3. Mechanical properties of the porous titanium body
We assessed the porous body’s mechanical properties in accordance with the International Organization for Standardization (ISO 13314:2011[E]), using porous cylinders (diameter 12 mm, length 12 mm) for the mechanical testing. We performed compressive testing using a universal material testing machine (Model EHF-LV020K1-010; Shimadzu Corp., Kyoto, Japan) at a crosshead speed of 1 mm/min. We calculated the elastic modulus from the slope of the compressive stress–strain curve in the linear elastic region. We determined the compressive yield strength from the stress–strain curve using the 0.2% offset method, resulting in an elastic modulus of 648.9 (±standard deviation [SD]=9.3) MPa and a compressive yield strength of 42 (±SD=1.4) MPa.
4. Bioactive surface treatment
The customized implants were treated chemically and thermally to provide a bioactive surface, as previously described [15]. Briefly, we immersed the fabricated porous titanium bodies in 5 M aqueous NaOH solution at 60°C for 24 hours, 0.5 mM HCl at 40°C for 24 hours, and ultrapure water at 40°C for 24 hours; we then heat treated them at 600°C for 1 hour. After the surface treatment, the whole porous surface was uniformly changed to a bioactive thin titanium oxide layer, approximately 1 mm thick, with sub-micron-sized pores.
5. In vitro evaluation of implant bioactivity
We confirmed the implants’ in vitro apatite-forming ability by soaking the samples for 3 days in acellular simulated body fluid (SBF) [16]. The porous body walls were completely covered with apatite within 3 days of soaking in SBF, rendering the whole surface of the implant bioactive with chemical and thermal treatments (Fig. 3). We sterilized the implants by exposing them to 25 kGy g of radiation before their surgical implantation.
6. Patients
The clinical trial enrolled four patients (three men, one woman; mean age, 62 years; range, 42–75 years) between January 2013 and October 2013 (Fig. 4). The inclusion criteria were symptomatic single or two-level cervical degenerative disc disease with compression of neural elements, refractory to appropriate conservative treatment for 3 or more months. We included patients with cervical spondylotic myelopathy or disc herniation and excluded those with ossification of the posterior longitudinal ligament. We excluded patients with multilevel (three or more) diseases, previous spinal surgery, history of trauma, general inflammatory disease, or severe comorbidities.
7. Surgical procedures
The two senior authors (S.F. and M.T.) performed all of the surgical procedures. Following a standard left anterolateral approach, the authors performed a complete discectomy and neural decompression. The bony endplate was preserved as much as possible to achieve perfect contact of the implant, the latter of which was inserted into the disc space. Stability was confirmed manually after the distractor was removed. Usually, the implants self-locate into the correct position, and moving the cage in any direction is not possible, thereby indicating excellent primary stability. The surgeons chose among three different cage heights in 1-mm steps depending on the intraoperative situation. After the surgery, we instructed the patients to remain in a soft collar for 4 weeks.
8. Radiological assessment
To assess bony union, we obtained lateral dynamic radiographs at 3, 6, and 12 months after the surgery. More than 3° of motion in flexion-extension was considered nonunion [17]. To evaluate bony union, we assessed coronal and sagittal multidetector CT reconstruction views at 3 and 12 months. Bone union was defined as complete when there was osseous continuity between the bony endplate and the implant in both the coronal and sagittal multidetector CT images. We defined nonunion as radiolucency >50% over the anteroposterior distance of the interface between the endplates and implants. Successful bony union was recorded when the previously mentioned radiological parameter assessments were complete. We calculated cage subsidence from the change in fused segment height (FSH), ascertained using the length of the lines drawn between the center of the cranial endplate of the cranial vertebrae and the center of the caudal end of the caudal vertebrae. We defined significant cage subsidence as an FSH change ≥3 mm.
The patients underwent magnetic resonance imaging (MRI) at 1 week and 12 months to assess neural decompression and adverse effects. Three independent experienced spinal surgeons, each with 10 years or more of experience, performed all the radiological assessments. Table 1 summarizes the preoperative clinical and radiological data.
9. Ethical considerations
The study was performed in accordance with the principles of the Declaration of Helsinki and the Good Clinical Practices and was registered in the University Hospital Medical Information Network Clinical Trials Registry. In all cases, the patient and his or her relatives were informed about the benefits and the risks of the implant. Written informed consent was obtained from all patients and/or their relatives, in accordance with protocols approved by the Institutional Ethics Committee of Kyoto University (C675).
10. Clinical assessment
We preoperatively and postoperatively examined the Japanese Orthopaedic Association (JOA) score and its recovery rate (recovery rate=postoperative score−preoperative score/17 [full score]−preoperative score×100%). We assessed the self-reported neck pain Visual Analog Scale scores before the operation and at 7 and 14 days after the operation.
11. Endpoints
The primary endpoints of this clinical trial were improvement in the clinical results and no implant-related adverse effects. The secondary endpoints were a lack of need for an autologous iliac crest bone graft and bony union without implant subsidence.
12. Clinical results
In all four patients, the preoperative myelopathy and radicular symptoms were resolved immediately after surgery, and the postoperative Visual Analog Scale scores decreased rapidly. There were no surgery-related complications.
13. Radiological results
The mean FSH was 31.06±3.41 mm preoperatively, 32.06±3.83 mm immediately after surgery, 31.7±3.85 mm 3 months after surgery, and 31.46±3.93 mm 1 year after surgery. The FSH values remained steady during the follow-up period, indicating no implant subsidence (Fig. 5). We had complete radiological parameters for all but one patient and considered complete bony union achieved in all but one patient 12 months after surgery. Table 2 summarizes the postoperative clinical and radiological results.
14. Illustrative case
A 74-year-old man with progressive myelopathy had a preoperative JOA score of 11.5. Severe spinal cord compression and an intramedullary T2 high intensity zone were revealed at the C4–5 level on MRI. The patient underwent ACDF with an individualized implant without bone graft according to the clinical trial protocol. The patient’s postoperative course was uneventful, and his JOA score improved to 16.5, with a recovery rate of 90.9% 1 year after surgery. The patient achieved radiological bony union within 1 year (Fig. 6).
Discussion
We successfully treated four patients with degenerative cervical disease using customized bioactive porous titanium implants without bone grafts. The clinical scores improved for all patients, and all but one achieved radiological bony union by the 5-year follow-up.
McGilvray et al. [18] evaluated the bone ingrowth potential of three types of cages in an ovine lumbar fusion model, comparing a polyether ether ketone (PEEK) cage, a titanium-coated PEEK cage, and a 3D-printed porous titanium cage. The results indicated that the 3D-printed porous titanium cage was superior, both biomechanically and histologically [18]. Several case reports have reported successful results following customized implants. Xu et al. [19] reported that Ewing sarcoma of the cervical spine was successfully removed and reconstructed using a customized artificial vertebral body. The implant was manufactured using electron beam technology, and radiological examinations revealed good osseointegration 1 year after surgery [19]. Spetzger et al. [20] reported an ACDF treatment using a customized cervical fusion cage whereby the bearing surface of a commercially-available titanium cage was customized using patient data. Although the authors described the manufacturing process and reported excellent primary stability, they did not evaluate the clinical results (such as bony union, cage subsidence, and neurological recovery). Their cage had a similar shape to ours; however, the customization was limited to the bearing surface. Moreover, the surface of other implants was not bioactive. Our implant is not only produced by AM technology using specific patient data but also provides surface bioactivity, providing both mechanical interlocking and biological bonding to the surrounding bone.
When manufacturing customized implants, the most time-consuming process is the design. Although experienced spine surgeons designed the implant for our clinical trial, which reduced the time required, the mean manufacturing period was 2–3 weeks. In the clinical practice, well-trained dedicated engineers should design the implants, and the simulated design should be shown to clinicians. Establishing an efficient processing chain is key to producing customized implants. A manufacturing period of 2–3 weeks might be acceptable for managing degenerative diseases. If we are to apply our technology to trauma surgery, however, reducing the manufacturing period is vital.
This trial had several limitations, the first of which was its small sample size and single-center nature. Second, the application of customized implants was limited to single or two-level ACDF. To establish the safety and efficacy of customized implants, further studies are needed to evaluate the application of this technology to other orthopedic fields.
In conclusion, although it is difficult to draw conclusions from this small clinical trial, our individualized bioactive porous titanium implant is a safe and promising implant for stand-alone ACDF, although further studies are needed to improve the design and manufacturing processes.
Notes
No potential conflict of interest relevant to this article was reported.
Funding
This study was supported by a Grant in Aid for Scientific Research from the Japan Society for the Promotion of Science (no., 19200039).
Author Contributions
Conception and design: SF, MT, TN, TM, TK, KS, SM; data acquisition: SF, MT, TM, KS, SM; analysis of data: SF, MT, TM, KS, SM; drafting of the manuscript: SF, MT; critical revision: SF, MT, TN, TM, TK, KS, SM; obtaining funding: SF, TN; supervision: TN, TK, SM; and approval of the final manuscript: all authors.