Correlation between the Computed Tomography Values of the Screw Path and Pedicle Screw Pullout Strength: An Experimental Study in Porcine Vertebrae

Article information

Asian Spine J. 2020;14(3):265-272
Publication date (electronic) : 2020 January 8
doi : https://doi.org/10.31616/asj.2019.0170
1Department of Orthopaedic Surgery, Kansai Medical University, Hirakata, Japan
2Department of Orthopaedic Surgery, Kansai Medical University Medical Center, Moriguchi, Japan
Corresponding author: Atsushi Ikeura Department of Orthopaedic Surgery, Kansai Medical University, 2-5-1 Shinmachi, Hirakata, Osaka 573-1010 Japan Tel: +81-72-804-0101, Fax: +81-72-804-2439, E-mail: a_ikeike55@mac.com
Received 2019 May 23; Revised 2019 July 19; Accepted 2019 August 19.

Abstract

Study Design

Biomechanical study.

Purpose

To assess the correlation between the computed tomography (CT) values of the pedicle screw path and screw pull-out strength.

Overview of Literature

The correlation between pedicle screw pull-out strength and bone mineral density has been well established. In addition, several reports have demonstrated a correlation between bone mineral density and CT values. However, no previous biomechanical studies investigated the correlation between CT values and pedicle screw pull-out strength.

Methods

Sixty fresh-frozen lumbar vertebrae from 6-month-old pigs were used. Before screw insertion, the CT values of the screw path were obtained for each sample. Specimens were then randomly divided into three equal groups. Each group had one of three pedicle screws inserted: 4.0-mm LEGACY (4.0-LEG), 4.5-mm LEGACY (4.5-LEG), or 4.5-mm SOLERA (4.5-SOL) (all from Medtronic Sofamor Danek Inc., Memphis, TN, USA). Each screw had a consistent 30-mm thread length. Axial pull-out testing was performed at a rate of 1.0 mm/min. Correlations between the CT values and pedicle screw pull-out strength were evaluated using Pearson’s correlation coefficient analysis.

Results

The correlation coefficients between the CT values of the screw path and pedicle screw pull-out strength for the 4.0-LEG, 4.5-LEG, and 4.5-SOL groups were 0.836 (p<0.001), 0.780 (p<0.001), and 0.873 (p<0.001), respectively. Greater CT values were associated with greater screw pull-out strength.

Conclusions

The CT values of the screw path were strongly positively correlated with pedicle screw pull-out strength, regardless of the screw type and diameter, suggesting that the CT values could be clinically useful for predicting pedicle screw pull-out strength.

Introduction

Recent advances in spinal surgical techniques have led to an increase in the proportion of patients undergoing instrumented spinal surgery [1]. Instrumented spinal surgery using pedicle screws is one of the most common methods for correcting spinal instability. However, the relatively poor bone quality in older adult patients’ results in decreased pedicle screw anchorage, potentially resulting in failure at the bone–screw interface [2,3]. The pedicle screw loosening rate in patients with osteoporosis reportedly ranges from 0% to 62.8% [4]. When pedicle screw fixation failure occurs, revision surgery may be required; therefore, achieving adequate pedicle screw anchorage is a major concern.

Spinal stability after treatment with instrumentation depends mainly on the stability of the pedicle screws and the vertebral bodies [3]. The stability of pedicle screws is affected by many factors, namely, screw design [5], pedicle morphology [6], and vertebral bone mineral density (BMD). Several studies have demonstrated a strong correlation between pedicle screw pull-out strength and the BMD of human vertebral bodies [3,7], and a recent radiological study reported a correlation between the pedicle screw pull-out strength and the pedicle BMD measured using dual-energy computed tomography (CT) [8]. Currently, BMD is assessed mainly with dual-energy X-ray absorptiometry (DXA), which is the only diagnostic method included in the World Health Organization’s definition of osteoporosis [9]. Furthermore, DXA is reportedly a powerful predictor of bone degradation and fracture risk [10]; however, this technique has limitations because it can be used only in the lumbar spine and the proximal femur, and is inaccurate in patients with lumbar scoliosis or degenerative changes[11].

There is current interest in using CT values to assess BMD. Some studies have reported a correlation between CT values and BMD, and proposed that CT values are a suitable alternative to DXA when measuring BMD [12,13]. Therefore, we hypothesized that the CT values of the screw path may be correlated with pedicle screw pull-out strength. However, to our knowledge, based on an English-language literature review, no previous biomechanical studies have investigated this relationship. Thus, we performed an axial pull-out test of pedicle screws from the lumbar vertebrae of porcine specimens. The aim of the present study was to assess the correlation between the CT values of the screw path and the pedicle screw pull-out strength, and to confirm the validity of using the CT values of the screw path to predict pedicle screw pull-out strength.

Materials and Methods

1. Ethical approval and consent to participate

The present study did not involve human participants or experimental animals, so ethical approval was not required.

2. Screw types

Three types of compliance engineering-certified monoaxial pedicle screws (Medtronic Sofamor Danek USA Inc., Memphis, TN, USA) were used: 4.0-mm LEGACY (4.0-LEG; diameter, 4.0 mm; length, 30 mm; thread pitch, 2.75 mm), 4.5-mm LEGACY (4.5-LEG; diameter, 4.5 mm; length, 30 mm; thread pitch, 2.75 mm), and 4.5-mm SOLERA (4.5-SOL; diameter, 4.5 mm; length, 30 mm; thread pitch, 2.0 mm in the cortical zone and 4 mm in the cancellous zone) (Fig. 1).

Fig. 1.

Photographs showing the screw types used in the present study. (A) 4.0-LEG (LEGACY; diameter, 4.0 mm; length, 30 mm), (B) 4.5-LEG (LEGACY; diameter, 4.5 mm; length, 30 mm), and (C) 4.5-SOL (SOLERA; diameter, 4.5 mm; length, 30 mm).

3. Specimen preparation

A total of 60 fresh-frozen lumbar vertebrae from various lumbar levels (cut into halves through the spinous processes) were obtained from 6-month-old pigs (Tokyo Shibaurazouki Co. Ltd., Tokyo, Japan) that had been slaughtered domestically for edible use (and so were not experimental animals). Porcine specimens were chosen because of their easy availability and use in previous studies [14,15]. In preparation for biomechanical testing, the specimens were thawed at room temperature; carefully cleaned of muscle, ligament, and connective tissue; and separated into individual vertebrae. These 60 vertebrae were randomly divided into three equal groups, one group for each screw type. Just prior to testing, the discs and cartilaginous endplates were resected. Screws were then inserted perpendicularly to the resection plane into pilot holes that had been predrilled with a 2.0-mm drill bit, which was smaller than the inner diameter of the screw tip. Pedicle screws were inserted directly into the vertebrae in the cranial–caudal direction, rather than transpedicularly, to eliminate the effects of the cortical bone surrounding the pedicle and the pedicular morphology. Tapping was not performed prior to screw insertion. All screws were inserted through a 2-mmthick washer with a consistent 30-mm thread length, and all screws were inserted by a single researcher to eliminate technical bias (Fig. 2A).

Fig. 2.

Images of the porcine vertebrae. (A) Photograph of a 30-mmlong screw inserted through a 2-mm-thick washer. (B) Computed tomographic image showing the 10×30-mm rectangular region of interest (enclosed by the black lines); cortical margins were excluded to prevent volume averaging.

4. Radiographic analysis

Before screw insertion, we measured the CT (Aquilion 16SX; Canon Medical Systems Corp., Otawara, Japan) value of the screw path in each specimen. The CT parameters included a slice thickness of 1.0 mm, tube voltage of 120 kVp, tube current of 300 mA, and a bone reconstruction algorithm (window width/window level: 3,000/400). A picture archiving and communication system (EV Insite R; PSP Corp., Tokyo, Japan) was used to calculate the average CT value for a region of interest (ROI) that was confined to the trabecular area. ROIs were measured using sagittal and coronal images. For each measurement, a 10×30-mm rectangular ROI was drawn without using any special device; the cortical margins were excluded to prevent volume averaging (Fig. 2B). The CT values from the two planes were averaged to obtain a mean value, and ROIs were selected by two researchers to evaluate the intra- and inter-observer variabilities of the CT values using the interclass correlation coefficient (ICC) values.

5. Biomechanical testing and analysis

To ensure that the screw tracts were parallel to the pull-out axis, specimens were placed in a custom-made stainless steel device that enabled capturing of the screw head and specimen for biomechanical testing (Fig. 3AC). This device had a rectangular shape with a screw hole at the bottom for rigid fixation within the lower part of the testing machine (Fig. 3A). On the upper side, the device had a rectangular hole through which only the pedicle screw head could pass, allowing the pull-out force applied by the machine to be transmitted directly to the specimen (Fig. 3B). The screw heads were attached to the testing machine, and a constant pull-out displacement rate of 1 mm/min was applied in line with the screw axis.

Fig. 3.

Biomechanical testing. (A) The autograph used to determine pull-out strength (N) at a displacement rate of 1.0 mm/min. (B) A screw head attached to the testing machine. (C) The bone–screw complex fixed inside the metal box. (D) Typical load–displacement curve for screw axial pull-out.

The maximum pull-out strengths of the pedicle screws were determined using a mechanical testing machine (Autograph AG-I; Shimadzu Corp., Kyoto, Japan). The load cell capacity of this system was 5,000 N, and the screw–bone complex was pulled until failure occurred. Data collected using TrapeziumX ver. 1.3.0 (Shimadzu Corp.), were used to calculate the ultimate failure load of each screw–bone complex. The maximum pull-out strength (N) was determined as the maximum point on the strength–displacement curve where failure occurred (Fig. 3D). Finally, we assessed the correlations between the CT values of the screw paths and the pull-out strengths of the three pedicle screws.

6. Statistical analysis

Intra- and inter-observer reliability calculations were performed using the ICC values, and were reported as scores between 0 and 1 (0, no agreement; 1, perfect agreement) [16]. ICC values of 0.75–0.89 were defined as indicating good reliability, while values of 0.90–0.99 were defined as excellent reliability. Differences between the CT values and pedicle screw pull-out strengths in each group were evaluated with one-way analysis of variance followed by a Tukey multiple comparison test. Correlations between the CT values of the screw path and the pedicle screw pull-out strengths were evaluated using Pearson’s correlation coefficient. Analyses were performed withJMP software ver. 10 (SAS Institute Japan Ltd., Tokyo, Japan). Two-sided p-values <0.05 were considered to indicate statistically significant differences.

Results

1. Computed tomography values and pull-out strengths

As shown in Table 1, the CT values of the screw paths did not differ significantly among the three groups (4.0-LEG versus 4.5-LEG, p=0.10; 4.0-LEG versus 4.5-SOL, p=0.78; 4.5-LEG versus 4.5-SOL, p=0.32). However, the 4.5-LEG and 4.5-SOL screws had significantly greater pull-out strengths than the 4.0-LEG screw (4.0-LEG versus 4.5-LEG, p=0.01; 4.0-LEG versus 4.5-SOL, p=0.03; 4.5-LEG versus 4.5-SOL, p=0.87).

Computed tomography values and pull-out strengths for each screw type

2. Correlation between the computed tomography value of the screw path and the pull-out strength

The correlation between the CT value of the screw path and pedicle screw pull-out strength was evaluated separately for each screw type (Fig. 4). The correlation coefficients (r) for the 4.0-LEG, 4.5-LEG, and 4.5-SOL screws were 0.836, 0.780, and 0.873, respectively. The CT values of the screw path were significantly positively correlated with pull-out strength for all three pedicle screws (p<0.001). In all groups, greater CT values were associated with greater screw pull-out strength.

Fig. 4.

Relationships between the CT values and pull-out strengths (N) of (A) 4.0-LEG (LEGACY; diameter, 4.0 mm; length, 30 mm), (B) 4.5-LEG (LEGACY; diameter, 4.5 mm; length, 30 mm), and (C) 4.5-SOL (SOLERA; diameter, 4.5 mm; length, 30 mm) screws. The CT values were significantly positively correlated with the pull-out strengths (N) of all pedicle screws (p<0.001). CT, computed tomography; HU, Hounsfield units.

3. Reliability

The ICCs for intra- and inter-observer variability were 0.962 and 0.957, respectively, indicating excellent reliability of the CT value measurements.

Discussion

Correlations between the BMD of human vertebral bodies and the pedicle screw pull-out strength, and between the BMD and CT values have been proven [3,7,12,13]; however, the association between pedicle screw pull-out strength and CT values has not yet been discussed. To our knowledge, the current study is the first to evaluate the association between pedicle screw pull-out strength and CT values, and we found a strong correlation between the CT values of the screw path and the pedicle screw pull-out strength, regardless of the screw type and diameter (Fig. 4). Our results indicate that the CT values of the screw path may be useful in predicting pedicle screw pull-out strength, which may help spine surgeons develop better treatments.

Some studies have reported a strong correlation between implant stability and CT values. In dentistry, the correlation between CT values and implant stability has been shown in vitro and in vivo, and normal CT values have been reported for specific regions of the mandible and maxilla [17,18]. In orthopedic surgery, Aichmair et al. [19] demonstrated a trend towards a greater pull-out strength with a greater vertebral BMD measured using CT values. However, their research had two drawbacks: BMD was evaluated in the ellipsoidal regions of the vertebral body (not the screw path), and the pull-out strength was compared among only three subgroups divided according to CT values [19]. Therefore, to further clarify the effect of CT values on pull-out strength, we investigated the correlation between the CT values of the screw path and the pull-out strength of the pedicle screw. Matsukawa et al. [20] used the CT value of the screw path as a BMD measurement and showed that the CT value correlated with the pedicle screw insertional torque. However, the authors also stated that insertional torque, which represents the holding power in the axial direction, closely approximates but is not identical to the pull-out strength [20]. Furthermore, the relationship between insertional torque and pull-out strength remains controversial [21,22], and screw pull-out occurs occasionally, clinically [2,23]. Thus, in the current study, we focused on the relationship between the CT values of the screw path and the pedicle screw pull-out strength. In addition to the strong positive correlation between the CT values of the screw path and pedicle screw pull-out strength in our study, the correlation coefficients (0.836, 0.780, and 0.873 for the 4.0-LEG, 4.5-LEG, and 4.5-SOL groups, respectively) were comparable with the correlation coefficient reported in a previous study (p=0.80) demonstrating a correlation between the pedicle screw pull-out strength and the pedicle BMD measured using dual-energy CT [8].

CT values, expressed in Hounsfield units (HU), are standard measurements representing the relative densities of body tissues according to a calibrated grayscale based on the values for air (−1,000 HU) and water (0 HU). Schreiber et al. [12] reported that participants with normal bone density, osteopenia, and osteoporosis had mean lumbar CT values of 133.0 HU, 100.8 HU, and 78.5 HU, respectively. The authors also showed correlations between CT values and the T-scores cited in the World Health Organization’s guidelines for diagnosing osteoporosis, and concluded that CT values can be used for fracture risk assessment, and for the diagnosis and early initiation of the treatment of osteoporosis [12]. Additionally, microarchitectural studies have reported that CT values can approximate microarchitectural parameters such as bone surface density [24], and that bone surface density is the most effective tool to predict pedicle screw pull-out strength [25]. In addition, CT values can be measured easily in any form by drawing an ROI on the captured CT image (Fig. 2B). However, the reproducibility of ROI selection is an important concern. In the present study, two spine surgeons manually placed ROIs on the images of 60 porcine lumbar vertebrae. The ICCs for intra- and interobserver variabilities were 0.962 and 0.957, respectively, which were similar to values reported in previous studies [12,13,20] showing excellent reliability of CT value measurements in the spine region. Although there is room for improvement via the development of software applications to automatically select optimal ROIs, CT values can be measured simply and accurately without a specialized device.

CT values have advantages over DXA for determining screw stability. A study published in 2009 reported that only 44% of spine surgeons use DXA to preoperatively assess BMD [26]. DXA incurs additional cost and radiation exposure (a typical total-body DXA examination exposes patients to 0.96 mSv) [27], and the technique can be used only to assess the lumbar spine and the proximal femur. In contrast, CT scans are always performed preoperatively to evaluate the morphology of the vertebral body and pedicle (especially the pedicular diameter) to determine the appropriate screw size, and can be obtained for any vertebra in which screw placement is planned. Therefore, we considered CT values to be more suitable for estimating pedicle screw pull-out strength than DXA, particularly for clinical use.

In the present study, we excluded the cortical margins when selecting the ROIs. A previous cadaver study reported that the mean percentage of the cortical bone contacting the screw inserted in the vertebral pedicle was only 2.7%, and that the cortical bone of the vertebral pedicle was not an important contributing factor in the fixation of implants [28]. Although part of the pedicle screw pull-out strength may be attributed to its interface with the cortical margins of the pedicle, we considered that the association between the pedicle screw and cancellous bone was more critical for evaluating screw stability.

The present study had several limitations that should be noted. First, the BMD of porcine bone is greater than that of human bone [29]. We used porcine vertebrae because our cultural background makes it difficult to obtain cadaveric spinal specimens. However, many studies have documented the utility of porcine bone as a model [14,15]. Furthermore, porcine vertebrae are similar to human vertebrae in terms of their nutritional make-up, bone structure, and mineral metabolism [30], and the microarchitectural ratios of cortical bone and cancellous bone in porcine vertebrae are similar to those of human vertebrae [24]. Although a cadaveric study using human bones of varying quality would be ideal, the goal of the present study was to determine the relationship between the CT values of the screw path and pedicle screw pull-out strength, specifically. Second, there are currently no reference CT values for the vertebrae, and our results only demonstrate a correlation between the CT values of the screw path and pedicle screw pull-out strength. Ideally, reference values for clinical use should be determined in future studies.

Conclusions

We introduced a novel application of CT values as a simple measurement that was significantly positively correlated with pedicle screw pull-out strength. Our findings suggest that the CT values of the screw path can be used to predict pedicle screw pull-out strength, and may be a reasonable substitute for BMD measured by DXA. However, our findings must be confirmed in a larger cadaveric study.

Notes

No potential conflict of interest relevant to this article was reported.

Funding

This work was funded by a Ministry of Education, Culture, Sports, Science and Technology (MEXT)-supported program of the Strategic Research Foundation for private universities (grant no., S1201038).

Author Contributions

Atsushi Ikeura: conception and design, data acquisition, analysis of data, drafting of the manuscript, funding acquisition; Taketoshi Kushida: data acquisition, critical revision; Kenichi Oe: drafting of the manuscript, critical revision; Yoshihisa Kotani: critical revision; Muneharu Ando: critical revision; Takashi Adachi: administrative support; and Takanori Saito: supervision.

Acknowledgements

The authors acknowledge Medtronic (Medtronic Sofamor Danek, Memphis, TN, USA) for their donation of the pedicle screws used in the study. We also thank Kelly Zammit, BVSc, and Jane Charbonneau, DVM, from Edanz Editing, for editing a draft of this manuscript.

References

1. Gaines RW Jr. The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 2000;82:1458–76.
2. Esses SI, Sachs BL, Dreyzin V. Complications associated with the technique of pedicle screw fixation: a selected survey of ABS members. Spine (Phila Pa 1976) 1993;18:2231–9.
3. Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS 3rd, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976) 1994;19:2415–20.
4. Galbusera F, Volkheimer D, Reitmaier S, BergerRoscher N, Kienle A, Wilke HJ. Pedicle screw loosening: a clinically relevant complication? Eur Spine J 2015;24:1005–16.
5. Kim YY, Choi WS, Rhyu KW. Assessment of pedicle screw pull-out strength based on various screw designs and bone densities: an ex vivo biomechanical study. Spine J 2012;12:164–8.
6. Hirano T, Hasegawa K, Takahashi HE, et al. Structural characteristics of the pedicle and its role in screw stability. Spine (Phila Pa 1976) 1997;22:2504–10.
7. Soshi S, Shiba R, Kondo H, Murota K. An experimental study on transpedicular screw fixation in relation to osteoporosis of the lumbar spine. Spine (Phila Pa 1976) 1991;16:1335–41.
8. Wichmann JL, Booz C, Wesarg S, et al. Quantitative dual-energy CT for phantomless evaluation of cancellous bone mineral density of the vertebral pedicle: correlation with pedicle screw pull-out strength. Eur Radiol 2015;25:1714–20.
9. Lewiecki EM, Gordon CM, Baim S, et al. International Society for Clinical Densitometry 2007 adult and pediatric official positions. Bone 2008;43:1115–21.
10. Stone KL, Seeley DG, Lui LY, et al. BMD at multiple sites and risk of fracture of multiple types: long-term results from the Study of Osteoporotic Fractures. J Bone Miner Res 2003;18:1947–54.
11. Rand T, Seidl G, Kainberger F, et al. Impact of spinal degenerative changes on the evaluation of bone mineral density with dual energy X-ray absorptiometry (DXA). Calcif Tissue Int 1997;60:430–3.
12. Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am 2011;93:1057–63.
13. Lee S, Chung CK, Oh SH, Park SB. Correlation between bone mineral density measured by dual-energy X-ray absorptiometry and Hounsfield units measured by diagnostic CT in lumbar spine. J Korean Neurosurg Soc 2013;54:384–9.
14. Gurwitz GS, Dawson JM, McNamara MJ, Federspiel CF, Spengler DM. Biomechanical analysis of three surgical approaches for lumbar burst fractures using short-segment instrumentation. Spine (Phila Pa 1976) 1993;18:977–82.
15. Hasegawa K, Ikeda M, Washio T, Hara T. An experimental study of porcine lumbar segmental stiffness by the distraction-compression principle using a threaded interbody cage. J Spinal Disord 2000;13:247–52.
16. Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull 1979;86:420–8.
17. Turkyilmaz I, Tumer C, Ozbek EN, Tozum TF. Relations between the bone density values from computerized tomography, and implant stability parameters: a clinical study of 230 regular platform implants. J Clin Periodontol 2007;34:716–22.
18. Turkyilmaz I, Sennerby L, McGlumphy EA, Tozum TF. Biomechanical aspects of primary implant stability: a human cadaver study. Clin Implant Dent Relat Res 2009;11:113–9.
19. Aichmair A, Moser M, Bauer MR, et al. Pull-out strength of patient-specific template-guided vs. free-hand fluoroscopically controlled thoracolumbar pedicle screws: a biomechanical analysis of a randomized cadaveric study. Eur Spine J 2017;26:2865–72.
20. Matsukawa K, Abe Y, Yanai Y, Yato Y. Regional Hounsfield unit measurement of screw trajectory for predicting pedicle screw fixation using cortical bone trajectory: a retrospective cohort study. Acta Neurochir (Wien) 2018;160:405–11.
21. Zdeblick TA, Kunz DN, Cooke ME, McCabe R. Pedicle screw pull-out strength: correlation with insertional torque. Spine (Phila Pa 1976) 1993;18:1673–6.
22. Inceoglu S, Ferrara L, McLain RF. Pedicle screw fixation strength: pull-out versus insertional torque. Spine J 2004;4:513–8.
23. Rivet DJ, Jeck D, Brennan J, Epstein A, Lauryssen C. Clinical outcomes and complications associated with pedicle screw fixation-augmented lumbar interbody fusion. J Neurosurg Spine 2004;1:261–6.
24. Teo JC, Si-Hoe KM, Keh JE, Teoh SH. Correlation of cancellous bone microarchitectural parameters from microCT to CT number and bone mechanical properties. Mater Sci Eng C Mater Biol Appl 2007;27:333–9.
25. Nakashima D, Ishii K, Nishiwaki Y, et al. Quantitative CT-based bone strength parameters for the prediction of novel spinal implant stability using resonance frequency analysis: a cadaveric study involving experimental micro-CT and clinical multislice CT. Eur Radiol Exp 2019;3:1.
26. Dipaola CP, Bible JE, Biswas D, Dipaola M, Grauer JN, Rechtine GR. Survey of spine surgeons on attitudes regarding osteoporosis and osteomalacia screening and treatment for fractures, fusion surgery, and pseudoarthrosis. Spine J 2009;9:537–44.
27. Xia Y, Ergun DL, Wacker WK, Wang X, Davis CE, Kaul S. Relationship between dual-energy X-ray absorptiometry volumetric assessment and X-ray computed tomography-derived single-slice measurement of visceral fat. J Clin Densitom 2014;17:78–83.
28. Defino HL, Vendrame JR. Role of cortical and cancellous bone of the vertebral pedicle in implant fixation. Eur Spine J 2001;10:325–33.
29. Smit TH. The use of a quadruped as an in vivo model for the study of the spine: biomechanical considerations. Eur Spine J 2002;11:137–44.
30. Lee CY, Chan SH, Lai HY, Lee ST. A method to develop an in vitro osteoporosis model of porcine vertebrae: histological and biomechanical study. J Neurosurg Spine 2011;14:789–98.

Article information Continued

Fig. 1.

Photographs showing the screw types used in the present study. (A) 4.0-LEG (LEGACY; diameter, 4.0 mm; length, 30 mm), (B) 4.5-LEG (LEGACY; diameter, 4.5 mm; length, 30 mm), and (C) 4.5-SOL (SOLERA; diameter, 4.5 mm; length, 30 mm).

Fig. 2.

Images of the porcine vertebrae. (A) Photograph of a 30-mmlong screw inserted through a 2-mm-thick washer. (B) Computed tomographic image showing the 10×30-mm rectangular region of interest (enclosed by the black lines); cortical margins were excluded to prevent volume averaging.

Fig. 3.

Biomechanical testing. (A) The autograph used to determine pull-out strength (N) at a displacement rate of 1.0 mm/min. (B) A screw head attached to the testing machine. (C) The bone–screw complex fixed inside the metal box. (D) Typical load–displacement curve for screw axial pull-out.

Fig. 4.

Relationships between the CT values and pull-out strengths (N) of (A) 4.0-LEG (LEGACY; diameter, 4.0 mm; length, 30 mm), (B) 4.5-LEG (LEGACY; diameter, 4.5 mm; length, 30 mm), and (C) 4.5-SOL (SOLERA; diameter, 4.5 mm; length, 30 mm) screws. The CT values were significantly positively correlated with the pull-out strengths (N) of all pedicle screws (p<0.001). CT, computed tomography; HU, Hounsfield units.

Table 1.

Computed tomography values and pull-out strengths for each screw type

Pedicle screw CT valuea) (HU) Pull-out strengtha) (N)
4.0-LEG 412.9±107.5 757.7±379.7
4.5-LEG 470.1±82.0 1157.0±275.0b)
4.5-SOL 432.5±96.8 1085.9±532.8b)

Values are presented as mean±standard deviation. 4.0-LEG: LEGACY; diameter, 4.0 mm; length, 30 mm; thread pitch, 2.75 mm; 4.5-LEG: LEGACY; diameter, 4.5 mm; length, 30 mm; thread pitch, 2.75 mm; 4.5-SOL: SOLERA; diameter, 4.5 mm; length, 30 mm; thread pitch, 2.0 mm in the cortical zone and 4 mm in the cancellous zone.

HU, Hounsfield units.

a)

Groups were compared using one-way analysis of variance followed by the Tukey multiple comparison test.

b)

p<0.05 for 4.0-LEG vs. 4.5-LEG and 4.5-SOL.